The P9B multi-gas/multi-range, metal sealed, Mass Flow Meter (MFM) uses the latest in electronics enabling it to meet the most critical of process gas flow measurement requirements. The P9B Mass Flow Meter has an integrated Baratron® Capacitance Manometer, 0-100 psia Full Scale, pressure transducer that allows the user to measure and monitor line pressure. The pressure can be observed via the multi-directional LED display located on the top of the MFM.
The following options are available for P9B High Performance Mass Flow Meters
Ordering Code Example: P9B013502C60030
Configuration Option |
Option Code |
---|---|
P9B Mass Flow Meter | P9B |
Gas (Per Semi Standard E52-0703) |
|
Nitrogen (N2) | 013 |
Ammonia (NH3) | 029 |
Sulfur Hexafluoride (SF6) | 110 |
Flow Range Full Scale |
|
5 sccm | 500 |
10 sccm | 101 |
20 sccm | 201 |
50 sccm | 501 |
100 sccm | 102 |
200 sccm | 202 |
500 sccm | 502 |
1000 sccm | 103 |
2000 sccm | 203 |
5000 sccm | 503 |
10000 sccm | 104 |
20000 sccm | 204 |
30000 sccm | 304 |
50000 sccm | 504 |
Fittings (compatible with) |
|
Swagelok 4 VCR | R |
C-seal (1.125 in.) | C |
W-seal (1.125 in.) | H |
W-seal (1.15 in.) | F |
Connector |
|
DeviceNet | 6 |
RS485 (uses 9 pin connector) | 5 |
15 pin D (Analog I/O) | B |
9 pin D (Analog I/O) | A |
Valve/Device Type |
|
No Valve (MFM) | 0 |
Reserved |
|
Reserved | 0 |
Firmware |
|
EtherCAT | 10 |
All Other I/O MKS will ship firmware revision current to date |
30 |
The multi-gas/multi-range capability, along with tight performance specifications for accuracy, allow users to minimize inventory of high flow MFM part numbers. The multi-gas/multi-range feature (along with other custom controls) is accessed through the MFM embedded diagnostic interface, that requires no special software or hardware to operate. A standard Ethernet cable and JAVA-enabled HTML browser, widely available, are all the tools needed. The critical gas parameters for typical high flow rate gases are already stored on the device. Configuring the device is simply a matter of selecting the gas from a drop down menu and specifying the desired full scale flow range. The diagnostic interface also allows the user to perform routine device health checks, plot flow, and store operating data for off-line analysis.
Digital I/O | DeviceNet™ | EtherCAT® |
---|---|---|
Input Power Required | +11 to +25 VDC per DeviceNet specification (@ <2.0 watts) | +24 VDC (<5 watts) |
Connector | 5 pin microconnnector (DeviceNet) | 2 x RJ-45 (comm.) male, M8 male, 5 pin (power) |
Data Rate Switch/Selection | 4 positions: 125, 250, 500K (Default), PGM (programmable over the network) | No switch |
Data Rate/Network Length | Data rate (user selectable) 125 Kbps, 500 meters (1,640 feet) 250 Kbps, 250 meters (820 feet) 500 Kbps, 100 meters (328 feet) |
100 Mbps |
MAC ID Switches/Addresses | 2 switches, 10 positions; 0,0 to 6,3 are hardware ID numbers 7,0 to 9,9 are software ID numbers (6,4 to 6,9 are unused and, if selected will default to hardware ID number 6.3) |
3 switches, 16 positions |
Network Size | Up to 64 nodes | Up to 4095 nodes |
Network Topology | Linear (trunkline/dropline) power and signal on same network cable | n/a |
Visual Indicators | LED network status (green/red) LED module status (green/red) Scrolling LED displays (MFC Type, Flow Full Scale, Gas Type, IP address, Instance Number (1 to 31) |
LED Power (green) LED Run (green) LED Error (red) LED Comm (green) |
Compliance | CE | CE |