Differential Pressure Transducer, 100 Torr, 0.5 in. Tube, 0-10 VDC, Bakeable

Technical Specs

  • Full Scale Range
    100 Torr
  • Type
    Pressure Transducer
  • Fitting Type
    0.5 in. OD tube
  • Sensor Type
    Industrial Capacitance Manometer
  • Electrical Connector
    AMP 14-pin ribbon
  • Measurement Type
    Differential
  • Accuracy
    0.5% of Reading (± temperature coefficients)
  • Output Signal
    0 - 10 VDC into >10 kOhm load
  • Exposed Materials
    Px side: Inconel
    Pr side: Inconel, Ceramic, Palladium, Stainless Steel, Glass
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Zero Temperature Coefficient
    0.02% of Full Scale/°C
  • Maximum Line Pressure
    40 psig (275 kPa)
  • Resolution
    0.01% of Full Scale
  • Maximum Overpressure
    120% of Full Scale or 20 psi (140 kPa), whichever is greater (for Full Scale ranges less than 1000 mmHg, the following restrictions apply: if high pressure is on the Px side of the sensor, the maximum differential overpressure allowable is 120% of Full Scale or 20 psi, whichever is greater, if high pressure is on the Pr side of the sensor, the maximum differential overpressure allowable is 120% of Full Scale Consult factory for higher Pr overpressure protection.)
  • Analog Output
    0 - 10 VDC
  • Volume
    Px side: 7.9 cc
    Pr side: 19.0 cc
  • Power Requirements
    ±15 VDC
  • Operating Temperature
    Sensor: 0 - 150°C (bakeable to 200°C with cable removed), Preamp: 0 - 50°C
  • Operating Temperature Range
    Sensor: 0 - 150°C (bakeable to 200°C with cable removed), Preamp: 0 - 50°C
  • Overpressure Limit
    140 kPa

Need help?

Contact an Applications Specialist by sending us an email