Compare Model Drawings, CAD & Specs Full Scale Range Second Range Gain Ratio Fitting Type Availability Price
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 10 & 100 Torr, 45°C, 8 VCR-F, DB15, 0.12%/Rdg Accuracy, 2 Trip Points
$2,850
3 Weeks
100 Torr 10 to 1 8 VCR female
3 Weeks
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 50 & 100 Torr, 45°C, 8 VCR-F, DB15, 0.12%/Rdg Accuracy, 2 Trip Points
$2,850
3 Weeks
100 Torr 2 to 1 8 VCR female
3 Weeks
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 100 & 1,000 Torr, 45°C, 8 VCR-F, DB15, 0.12%/Rdg Accuracy, 2 Trip Points
$2,850
3 Weeks
1000 Torr 10 to 1 8 VCR female
3 Weeks
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 100 & 1,000 Torr, 45°C, NW16 ISO-KF, DB15, 0.12%/Rdg Accuracy, 2 Trip Points
$2,850
3 Weeks
1000 Torr 10 to 1 NW16 ISO-KF
3 Weeks

Configuration Options

Description
  600-series heated absolute baratron capacitance manometer D24F Dual-Range Capacitance Manometer, Heated to 45°C, 1-1000 Torr, Trip Relays

Specifications

  • Type
    Pressure Transducer
  • Sensor Type
    Dual-Range Capacitance Manometer
  • Measurement Type
    Absolute
  • Sensor Temperature
    Controlled to 45°C
  • Full Scale Pressure Ranges
    10, 100, 1000 Torr
  • Analog Output
    0 to 10 VDC into > 10K Ω load
    Second Output: 0-10 VDC into > 10K Ω load over lower range
  • Analog Connector
    15-pin D-subminiture
  • Accuracy
    0.12% of Reading
  • Resolution
    0.001% of Full Scale
  • Zero Temperature Coefficient
    0.002% of Full Scale/°C
  • Span Temperature Coefficient
    0.02% of Reading/°C
  • Operating Temperature Range
    15° to 40°C
  • Volume
    6.3 cc
  • Warm-up Time
    2 hours
  • Overpressure Limit
    45 psia (310 kPa)
  • Materials Exposed to Process Gases
    Inconel®
  • Power Requirements
    ±15 VDC ±5% @ 0.40 Amps (max.)
  • Trip Point Level Signal
    0.01 to 10 VDC for each trip point (first output only)
  • Relay Contact Ratings
    SPDT, 1 Amp @ 30 VDC or 0.5 Amp @ 30 VAC resistive
  • Compliance
    CE

Features

Need help?

Contact an Applications Specialist by sending us an email