Sorry, this product is no longer offered for sale. The recommended replacement product is DA02B.1TBAS21CV0.

Heated Capacitance Manometer, 0.1 Torr, 100°C, 0.5 in. Tube, EtherCAT RJ45, DB15, Vertical Calibration

Technical Specs

  • Type
    Digital Pressure Transducer
  • Sensor Type
    Capacitance Manometer
  • Measurement Type
    Absolute
  • Sensor Temperature
    100°C
  • Full Scale Range
    0.1 Torr
  • Accuracy
    0.5% of Reading
  • Fitting Type
    0.5 in. OD tube
  • Communication Protocol
    EtherCAT
  • Analog Output
    0 - 10 VDC
  • Analog Connector
    15-pin D-sub
  • Digital Connector
    Two RJ45 EtherCAT jacks
  • Calibration
    Vertical
  • Zero Temperature Coefficient
    0.010% of Full Scale/°C
  • Span Temperature Coefficient
    0.02% of Reading/°C
  • Operating Temperature Range
    15° to 50°C
  • Volume
    6.3 cc
  • Warm-up Time
    4 hours
  • Overpressure Limit
    45 psia (310 kPa) or 120% of Full Scale, whichever is greater
  • Exposed Materials
    Inconel® and Incoloy® nickel alloys
  • Trip Points
    None
  • Compliance
    CE

Need help?

Contact an Applications Specialist by sending us an email