Pressure Transducer, Differential, 50 in. H2O, 4 VCR Male, 0-1 VDC, Bidirectional

Technical Specs

  • Type
    Pressure Transducer
  • Sensor Type
    Capacitance Manometer
  • Measurement Type
    Differential
  • Full Scale Range
    50 in. H2O
  • Analog Output
    0 - 1 VDC
  • Calibration
    Bidirectional
  • Fitting Type
    4 VCR male
  • Electrical Connector
    15-pin HD D-sub
  • Resolution
    0.01% of Full Scale
  • Accuracy
    0.50% of Full Scale
  • Zero Temperature Coefficient
    0.1% Full Scale/°C for standard accuracy specification
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Sensor Internal Volume
    Measurement Side: 1.4 cc
    Reference Side 9.0 cc
  • Operating Temperature Range
    0 - 50°C
  • Maximum Overpressure
    Measurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
  • Maximum Line Pressure
    40 psig (275 kPa)
  • Exposed Materials
    Measurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
  • Measurement Side Fitting
    4 VCR male
  • Reference Side Fitting
    4 VCR male
  • Mounting Bracket
    Standard
  • Trip Points
    2
  • Trip Relays
    Independently adjustable from -100% to +100% of Full Scale
  • Relay Contact Ratings
    DPDT, 1.0 A at 30 VDC or 0.3 amps at 120 VAC
  • Warm-up Time
    30 minutes
  • Power Requirements
    24 VDC
  • Compliance
    CE, SEMI S2-0706

Need help?

Contact an Applications Specialist by sending us an email