Pressure Transmitter, Differential, 1000 Torr, 4 VCR-F, 4-20 mA, Bidirectional, Terminal

Technical Specs

  • Type
    Vacuum/pressure Transducer
  • Measurement Type
    Differential
  • Full Scale Range
    1000 Torr
  • Analog Output
    4-20 mA
  • Calibration
    Bidirectional
  • Electrical Connector
    5-pin Terminal block adapter
  • Resolution
    0.01% of Full Scale
  • Accuracy
    0.50% of Full Scale
  • Zero Temperature Coefficient
    0.1% Full Scale/°C for standard accuracy specification
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Sensor Internal Volume
    Measurement Side: 1.4 cm3, Reference Side: 9.0 cm3
  • Operating Temperature Range
    0-50°C
  • Maximum Overpressure
    Measurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
  • Maximum Line Pressure
    40 psig (275 kPa)
  • Exposed Materials
    Measurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
  • Measurement Side Fitting
    4 VCR female
  • Reference Side Fitting
    4 VCR female
  • Mounting Bracket
    Standard
  • Trip Points
    1
  • Power Requirements
    24-32 VDC
  • Compliance
    CE, SEMI S2-0706

Need help?

Contact an Applications Specialist by sending us an email