$ EZ-EDS Version 3.3 Generated Electronic Data Sheet $ 176360 Rev B DeviceNet Electronic Data Sheet $ for 275553-##-#. $ $$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$ $ 11/5/10 JGL Modified for the new uP and software $ Edited from 014231-02 $ $$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$ $ 12/1/10 WMT - Modified for 275553 part numbers. $ $$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$ $ 12/8/10 WJG - Corrected trip point hysteresis range $ - Made Param14 read-only $ - Added and corrected IO_Info $ - Added enumeration for changing I/O $ $$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$$ [File] DescText = "GP275 Vacuum Gauge"; CreateDate = 11-05-2010; CreateTime = 10:10:00; ModDate = 12-13-2010; ModTime = 08:36:39; Revision = 2.0; $ EDS Revision 1.0 = Rev A, 2.0 = Rev B, etc. [Device] VendCode = 92; VendName = "Granville-Phillips Company"; ProdType = 28; ProdTypeStr = "Vacuum Pressure Gauge"; ProdCode = 2; MajRev = 2; MinRev = 10; ProdName = "GP275"; Catalog = "275553"; [IO_Info] Default = 0x0001; PollInfo = 0x0001, 1, 0; Input1 = 5, 0, 0x0001, "Pressure Output (REAL) + Status", 6, "20 04 24 01 30 03", "Gauge output (REAL) plus gauge status."; Input2 = 3, 0, 0x0001, "Pressure Output (UINT) + Status", 6, "20 04 24 02 30 03", "Gauge output (UNIT) plus gauge status."; Input3 = 4, 0, 0x0001, "Pressure Output (REAL)", 6, "20 04 24 04 30 03", "Gauge output (REAL)."; [ParamClass] MaxInst = 30; Descriptor = 0x000B; CfgAssembly = 0; [Params] Param1 = 0, 6,"20 30 24 01 30 0B", 0x0010, 0xC6, 1, "Device Status", "", "Represents the current state of the device.", 0,255,0, ,,,, ,,,, 0; Param2 = 0, 6,"20 30 24 01 30 0C", 0x0010, 0xD1, 1, "Exception Status", "", "Indication of alarms and warnings for this device. Bit 1 = Alarm, Bit 5 = Warning", 0,255,0, ,,,, ,,,, 0; Param3 = 0, 6,"20 30 24 01 30 0F", 0x0000, 0xC1, 1, "Alarm Enable", "", "Attribute used to enable or disable Alarm Exception bit.", 0,1,0, ,,,, ,,,, 0; Param4 = 0, 6,"20 30 24 01 30 10", 0x0000, 0xC1, 1, "Warning Enable", "", "Attribute used to enable or disable Warning Exception bit.", 0,1,0, ,,,, ,,,, 0; Param5 = 0, 6,"20 31 24 01 30 04", 0x0010, 0xC7, 2, "Data Units", "", "Determines the Units context of value and related attributes.", 769,777,769, ,,,, ,,,, 0; Enum5 = 769,"Torr", 776,"Millibar", 777,"Pascal"; Param6 = 0, 6,"20 31 24 01 30 05", 0x0030, 0xC1, 1, "Reading Valid", "", "Indicates that the Value attribute contains a valid value.", 0,1,1, ,,,, ,,,, 0; Param7 = 0, 6,"20 31 24 01 30 06", 0x0030, 0xCA, 4, "Value", "", "Analog input value.", 0,9.99e+9,760, ,,,, ,,,, 0; Param8 = 0, 6,"20 31 24 01 30 07", 0x0030, 0xD1, 1, "Status", "", "Alarm and Warning State of this object instance. Bits 0 and 1 indicate hardware failure, bits 2 and 3 indicate need for Zero or Gain adjustment.", 0,255,0, ,,,, ,,,, 0; Param9 = 0, 6,"20 31 24 01 30 0B", 0x0010, 0xC6, 1, "Offset-A Data type", "", "Determines the Data type of attribute Offset-A (real).", 202,202,202, ,,,, ,,,, 0; Param10 = 0, 6,"20 31 24 01 30 0C", 0x0010, 0xCA, 4, "Offset-A", "", "An amount added prior toGain to derive Value.", -10,0,0, ,,,, ,,,, 0; Param11 = 0, 6,"20 31 24 01 30 0D", 0x0010, 0xC6, 1, "Gain Data Type", "", "Determines the Data type of attribute Gain (real).", 202,202,202, ,,,, ,,,, 0; Param12 = 0, 6,"20 31 24 01 30 0E", 0x0010, 0xCA, 4, "Gain", "", "An amount scaled to derive Value.", -10,0,0, ,,,, ,,,, 0; Param13 = 0, 6,"20 31 24 01 30 0F", 0x0010, 0xCA, 4, "Unity gain Reference", "", "Specifies the value of the Gain attribute equivalent to a gain of 1.0", -0.82,0,0, ,,,, ,,,, 0; Param14 = 0, 6,"20 31 24 01 30 23", 0x0010, 0xC7, 2, "Gas Calibration Object Instance", "", "Indicates which Gas Calibration object instance is active for this object.", 0,0,0, ,,,, ,,,, 0; Param15 = 0, 6,"20 31 24 01 30 60", 0x0030, 0xD1, 1, "Status Extension", "", "Bit mapped byte providing additional status bits of an S-Analog Sensor instance", 0,0,0, ,,,, ,,,, 0; Param16 = 0, 6,"20 31 24 01 30 63", 0x0010, 0xC7, 2, "Subclass Number", "", "Indicates subclass extension of this object. This device is a Heat Transfer Vacuum Gauge.", 2,5,2, ,,,, ,,,, 0; Param17 = 0, 6,"20 34 24 01 30 03", 0x0010, 0xC7, 2, "Gas Standard number", "", "Gas Type Number, Nitrogen", 0,13,13, ,,,, ,,,, 0; Param18 = 0, 6,"20 35 24 01 30 05", 0x0000, 0xCA, 4, "Low Trip Point 1", "", "Defines the value below which a trip point condition will occur.", 0,9E+9,9E+9, ,,,, ,,,, 0; Param19 = 0, 6,"20 35 24 01 30 06", 0x0000, 0xC1, 1, "Low Trip Enable 1", "", "Enables the Low Trip Point setting.", 0,1,0, ,,,, ,,,, 0; Param20 = 0, 6,"20 35 24 01 30 07", 0x0030, 0xC1, 1, "Trip Point Status 1", "", "State of this object instance.", 0,0,0, ,,,, ,,,, 0; Param21 = 0, 6,"20 35 24 01 30 08", 0x0000, 0xC1, 1, "Trip Point Polarity 1", "", "Normal/Reverse polarity of Status attribute boolean.", 0,1,0, ,,,, ,,,, 0; Param22 = 0, 6,"20 35 24 01 30 09", 0x0030, 0xC6, 1, "Trip Point Override 1", "", "Specifies an Override status.", 0,0,0, ,,,, ,,,, 0; Param23 = 0, 6,"20 35 24 01 30 0A", 0x0000, 0xCA, 4, "Trip Point Hysteresis 1", "", "Specifies percentage above which a trip point will be cleared as pressure rises.", 0,10000,20, ,,,, ,,,, 0; Param24 = 0, 6,"20 35 24 02 30 05", 0x0000, 0xCA, 4, "Low Trip Point 2", "", "Defines the value below which a trip point condition will occur.", 0,9E+9,9E+9, ,,,, ,,,, 0; Param25 = 0, 6,"20 35 24 02 30 06", 0x0000, 0xC1, 1, "Low Trip Enable 2", "", "Enables the Low Trip Point setting.", 0,1,0, ,,,, ,,,, 0; Param26 = 0, 6,"20 35 24 02 30 07", 0x0030, 0xC1, 1, "Trip Point Status 2", "", "State of this object instance.", 0,0,0, ,,,, ,,,, 0; Param27 = 0, 6,"20 35 24 02 30 08", 0x0000, 0xC1, 1, "Trip Point Polarity 2", "", "Normal/Reverse polarity of Status attribute boolean.", 0,1,0, ,,,, ,,,, 0; Param28 = 0, 6,"20 35 24 02 30 09", 0x0030, 0xC6, 1, "Trip Point Override 2", "", "Specifies an Override status.", 0,0,0, ,,,, ,,,, 0; Param29 = 0, 6,"20 35 24 02 30 0A", 0x0000, 0xCA, 4, "Trip Point Hysteresis 2", "", "Specifies percentage above which a trip point will be cleared as pressure rises.", 0,10000,20, ,,,, ,,,, 0; Param30 = 0, 6,"20 04 24 00 30 65", 0x0000, 0xC6, 1, "Produced Instance Select", "", "Vendor specific attribute controlling Input I/O data type and size. 1 = two I/O bytes, 2 = three I/O bytes, 4 = four I/O bytes, 5 = five I/O bytes.", 1,5,5, ,,,, ,,,, 0; Enum30 = 1,"UINT Pressure", 2,"UINT Pressure + Status", 4,"REAL Pressure", 5,"REAL Pressure + Status";