  PBTS1B PBTS1B Portable Baratron® Transfer Standard

Specifications

  • Type
    Portable Baratron® Transfer Standard
  • Transfer Standard
    AA06A Absolute Sensor (NIST traceable)
  • Accuracy
    Standard 0.08% of Reading (±temp. coefficients)
    Optional (for 1-1000 range) 0.05% of Reading (±temp. coefficients)
  • Resolution
    1 x 10-6 of Full Scale
  • Fitting Type
    4 VCR® female
  • Operating Temperature
    15-40°C
  • Readout
    5½-digit LCD (670C)
  • Power Requirements
    110 VAC, 60 Hz; 220 VAC, 50 Hz; 100 VAC, 50/60 Hz; 3A
  • Warm-up Time
    4 hours. However, use of the internal 280 VA UPS will keep the transfer standard warmed up and ready for immediate use for at least 60 minutes.
  • Enclosure
    Metal wheeled caddy with extendible handle and tilt stand
  • Weight
    75 lbs (34 kg)
  • Configuration
    Removable transfer standard (10 ft./3m cable with 4 VCR female transducer fitting)

Features

PBTS1B Process Setpoint Verification

Typical process set point verification

The illustration above shows a typical process set point verification. The AA06A transfer standard of the PBTS1B is removed from its storage drawer and placed in close proximity to the process tool’s manometer port. This is facilitated by using a combination isolation valve such as the IDA™ valve. With the transfer standard connected to the IDA valve’s access port, it can be used to verify accuracy of the gauge and check system set point pressures.