Compare Model Drawings, CAD & Specs Full Scale Range Second Range Gain Ratio Calibration Fitting Type Availability Price
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 0.1 & 1 Torr, 45°C, 8 VCR-F, DB15, 0.12%/Rdg Accuracy, 2 Trip Points, Horizontal Calibration
1 Torr 10 to 1 Horizontal 8 VCR female
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 0.1 & 1 Torr, 45°C, NW16 ISO-KF, DB15, 0.12%/Rdg Accuracy, 2 Trip Points, Vertical Calibration
1 Torr 10 to 1 Vertical NW16 ISO-KF
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 1 & 10 Torr, 45°C, 8 VCR-F, DB15, 0.12%/Rdg Accuracy, 2 Trip Points, Vertical Calibration
$2,575
3 Weeks
10 Torr 10 to 1 Vertical 8 VCR female
3 Weeks
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 1 & 10 Torr, 45°C, NW16 ISO-KF, DB15, 0.12%/Rdg Accuracy, 2 Trip Points, Vertical Calibration
$2,575
3 Weeks
10 Torr 10 to 1 Vertical NW16 ISO-KF
3 Weeks
600-series heated absolute baratron capacitance manometer
Dual-range Heated Capacitance Manometer, 2 & 10 Torr, 45°C, NW16 ISO-KF, DB15, 0.12%/Rdg Accuracy, 2 Trip Points, Vertical Calibration
10 Torr 5 to 1 Vertical NW16 ISO-KF

Configuration Options

Description
  600-series heated absolute baratron capacitance manometer D27F Dual-Range Capacitance Manometer, Heated to 45°C, 1-1000 Torr

Specifications

  • Type
    Pressure Transducer
  • Sensor Type
    Dual-Range Capacitance Manometer
  • Measurement Type
    Absolute
  • Sensor Temperature
    Controlled to 45°C
  • Full Scale Pressure Ranges
    1, 10, 100, 1000 Torr
  • Analog Output
    0 to 10 VDC into > 10K Ω load
  • Analog Connector
    15-pin D-subminiture
  • Accuracy
    0.12% of Reading
  • Resolution
    0.001% of Full Scale
  • Zero Temperature Coefficient
    0.002% of Full Scale/°C
  • Span Temperature Coefficient
    0.02% of Reading/°C
  • Operating Temperature Range
    15° to 40°C
  • Volume
    6.3 cc
  • Warm-up Time
    2 hours
  • Overpressure Limit
    45 psia (310 kPa)
  • Materials Exposed to Process Gases
    Inconel®
  • Power Requirements
    ±15 VDC ±5% @ 0.25 Amps (max.)
  • Compliance
    CE

Features

Need help?

Contact an Applications Specialist bysending us an email