Compare Model Drawings, CAD & Specs Compatible with Interface Relays Units Electrical Plug Availability Price
370 stabil-ion vacuum gauge controller
$2,850
370 Stabil-Ion Gauge No 0 mbar North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$2,850
370 Stabil-Ion Gauge No 0 mbar North America 240 VAC
370 stabil-ion vacuum gauge controller
$2,850
370 Stabil-Ion Gauge No 0 mbar Europe 220 VAC
370 stabil-ion vacuum gauge controller
$2,850
370 Stabil-Ion Gauge No 0 mbar UK 240 VAC
370 stabil-ion vacuum gauge controller
$3,350
370 Stabil-Ion and 275 Convectron Gauges No 0 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$3,600
370 Stabil-Ion, 275 Convectron, and Capacitance Manometers No 0 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$3,300
370 Stabil-Ion Gauge RS232 0 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$3,825
370 Stabil-Ion and 275 Convectron Gauges RS232 0 Pascal North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$3,375
370 Stabil-Ion Gauge RS485 0 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$4,250
370 Stabil-Ion and 275 Convectron Gauges RS485 2 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$4,350
370 Stabil-Ion and 275 Convectron Gauges RS485 6 Torr North America & Japan 120 VAC
370 stabil-ion vacuum gauge controller
$3,375
370 Stabil-Ion Gauge IEEE 488 0 Torr North America & Japan 120 VAC

Specifications

  • Measurement Range
    UHV Stabil-Ion Gauge: 2 x 10-11 to 999 Torr; 3 x 10-11 to 1.33 x 103 mbar; 3 x 10-9 to 1.33 x 105 Pa
    Extended Range Stabil-Ion Gauge: 2 x 10-10 to 999 Torr;3 x 10-10 to 1.3 x 103 mbar;3 x 10-8 to 1.3 x 105 Pa
  • Display
    3 digits, plus exponent, green LED
  • Units
    orr, mbar, or Pa
  • Power Requirements
    90 to 130 Vac, or 180 to 250 VAC, 50/60 Hz, 220 W max
  • Communication Protocol
    RS-232, RS-485 or IEEE-488
  • Analog Output
    Stabil-Ion: 1 volt per decade, logarithmic, 0-10 VDC
    Convectron option: 1 volt per decade, logarithmic, 0-7 VDC
  • Set Point Contact Rating
    5 A at 250 Vac, 5 A at 30 Vac, resistive load
  • Set Point Configuration
    Single pole, double throw (SPDT)
  • Accuracy
    1 x 10-8 Torr to 1 x 10-4 Torr: ±4% of reading (N2)
  • Repeatability
    1 x 10-8 Torr to 1 x 10-4 Torr: ±3% of reading
  • Emission Current
    0.1 mA and 4.0 mA
  • Degas
    Electron bombardment, 40 W, 1 to 30 minutes (adjustable)
  • Operating Temperature
    0°C to 40°C
  • Storage Temperature
    -40°C to 70°C
  • Case Materials
    Aluminum extrusion, steel, plastic

Features

Need help?

Contact an Applications Specialist bysending us an email