Specifications
MultiGas 2030
More about the MultiGas 2030 here
|
Analyzer
|
|
Measurement Technique
|
FTIR Spectrometry
|
|
Gases and Vapors Measurable
|
Most molecules except for N2, H2, and O2
|
|
Ranges
|
Concentration setting between 10ppb and 100% full scale
|
|
FTIR
|
2102 Process FTIR
|
|
Spectral Resolution
|
0.5 128cm-1
|
|
Scan Speed
|
1 scan/sec @ 0.5cm-1
|
|
Scan Time
|
1-300 sec
|
|
Infrared Source
|
Silicon Carbide @ 1200°C
|
|
Reference Laser
|
Helium Neon (15798.2cm-1)
|
|
Detector
|
LN2 -cooled MCT; TE -cooled MCT
|
|
Purge Pressure
|
20 psig (1.5 bar) max.
|
|
Spectrometer Purge Flow
|
0.2 L/min of dry nitrogen or CO2 free clean dry air with dewpoints below -70°C
|
|
Optics Purge Flow
|
0.2 L/min of dry nitrogen or CO2 free clean dry air with dewpoints below -70°C
|
|
Pressure Transducer
|
MKS Baratron®
|
|
Purge Connection
|
Swagelok® quick connect
|
|
Computer Requirements
|
Desktop or notebook Intel Pentium® PC under Microsoft® Windows® XP; display (1024 x 768)
|
|
Recommended Computer
|
Intel Pentium III, 850 MHz, Microsoft® Windows® XP, 256 MB
|
|
Communications
|
RJ-45 cross-over Ethernet
|
|
Output Options
|
XML, analog, AK, others (please inquire)
|
|
Dimensions
|
17.5"W x 12.5"H x 25.5"D
|
|
Installation
|
19" rack mount chassis
|
|
Power
|
120 or 240 VAC, 50/60 Hz, 3 amps
|
|
Weight
|
110 lbs. (50 kg)
|
|
Sampling Parameters
|
|
Sample Temperature
|
Ambient to 200°C (calibration temperature dependant)
|
|
Sample Flow
|
0.2 10 L/min
|
|
Sample Pressure
|
0.01 4 atm (calibration pressure dependant)
|
|
Gas Cell
|
|
Construction
|
Nickel coated Al; Welded 316 stainless steel optional
|
|
Fittings
|
¼" Swagelok®, ¼" VCR®
|
|
Tubing
|
Heated ¼" stainless steel
|
|
Mirrors
|
Nickel plated aluminum substrate, with rugged gold coating
|
|
Windows
|
KBr; CaF2 (others available)
|
|
O-rings
|
Viton® (others available)
|
|
Detection Limits
|
|
Low-level detection limits for the 5.11 meter gas cell and a mercury-cadmium-telluride (MCT) detector at 0.5 cm-1 resolution for typical gases in the absence of interfering species are below:
|
|
Name (formula)
|
Lowest Detectable Limit with 20/20 Cell and 5 min Measurement
|
Lowest Detectable Limit with 20/20 Cell and 1 sec Measurement
|
|
Ammonia (NH3)
|
30ppb
|
0.5ppm
|
|
Carbon Dioxide (CO2)
|
12ppb
|
0.2ppm
|
|
Carbon Monoxide (CO)
|
72ppb
|
1.2ppm
|
|
Formaldehyde (H2CO)
|
36ppb
|
0.6ppm
|
|
Hydrogen Chloride (HCl)
|
87ppb
|
1.5ppm
|
|
Hydrofluoric Acid (HF)
|
12ppb
|
0.2ppm
|
|
Methane (CH4)
|
36ppb
|
0.6ppm
|
|
Nitrogen Dioxide (NO2)
|
24ppb
|
0.4ppm
|
|
Nitric Oxide (NO)
|
210ppb
|
3.6ppm
|
|
Nitrogen Trifluoride (NF3)
|
30ppb
|
0.5ppm
|
|
Silicon Tetrafluoride (SiF4)
|
10ppb
|
0.15ppm
|
|
Sulfur Dioxide (SO2)
|
36ppb
|
0.6ppm
|
|
Tetrafluoromethane (CF4)
|
2.5ppb
|
40ppb
|
|
Xylenes (C8H10)
|
60ppb
|
1.0ppm
|
Specifications are subject to change without notice.
Copyright MKS Instruments, Inc. unless noted otherwise.
© 2010 by MKS Instruments