Pressure/Vacuum Measurement

Baratron® Direct (Gas Independent) Pressure/Vacuum Capacitance Manometers (0.01-155,000 Torr)

HPS® Vacuum Transducers, Gauges, Sensors and Controllers (1x10-11 to 1500 Torr)

Pressure/Vacuum Switches

Specifications

872B Micro-Baratron®
High Pressure, Ultraclean, Non-Incendive Flow-through Pressure Transducers

More about the Micro-Baratron® 870B here
More about the Micro-Baratron® 872B here

870B & 872B Micro-Baratron®
Configuration Ultraclean, flow-through
Full Scale Ranges 1000 Torr to 3000 psia
Accuracy 1.0% of Reading
(including non-linearity, hysteresis, and non-repeatability)
Temperature Coefficients
Zero 0.02% of Full Scale/°C, 50 psia to 3000 psia
0.04% of Full Scale/°C, 1000 Torr
Span 0.04% of Reading/°C, 50 psia to 3000 psia
0.08% of Reading/°C, 1000 Torr
Ambient Operating Temperature 0° to 50°C (32° to 122°F)
Overpressure Limit 1.5 x Full Scale for ranges from 1000 Torr to 500 psi
1.2 x Full Scale for ranges from 501 to 3000 psi
(the pressure at which the transducer can be subjected without degradation of performance when returned to a normal operation pressure range)
Burst Pressure 10 x Full Scale for ranges from 1000 Torr to 500 psi
5 x Full Scale for ranges from 501 to 3000 psi
Materials Exposed to Gases Incoloy®
Surface Finish < 5µin Ra max.
Input Power Required
0 to 10 Volt output +13 VDC to +32 VDC @ 10 mA max.
0 to 5 Volt output +12 VDC to +32 VDC (regulated if below 13 VDC) @ 10 mA max.
2-wire 4-20 mA output +13 to +32 VDC excitation
Output
0 to 10 VDC into > 10K Ω load
0 to 5 VDC into > 10K Ω load
2-wire 4-20 mA output with +13 to +32 VDC at transducer terminals Typically into 0 to 900 W load (depending upon excitation)
Compliance CE compliant to EMC Directive 2004/108/EC2
(for CE compliance the mating connection must be properly grounded)
FM3611-1986 for Class I, Division 2, Groups A, B, C, and D
(4-20 mA output signal only)
Operating Environment Indoor operation standard, optional NEMA4-rated outdoor operation
(4-20 mA output signal only)
Electrical Connectors Male 9-pin or male 15-pin Type "D" at end of 9 inch flying lead,
Bendix® at end of 9 inch flying lead, or 6 foot or 10 foot flying leads
Fittings ¼ inch weld stub, 4 VCR® male, 4 VCR® female, surface mount

Specifications are subject to change without notice.
VCR® is a registered trademark of Swagelok® Company.
Copyright MKS Instruments, Inc. unless noted otherwise.


Need help?

Contact an Applications Specialist by sending an email to MKS Instruments, or call 978-975-2350.


MKS Instruments Logo

Global Headquarters
2 Tech Drive, Suite 201
Andover, MA 01810
Tel: 978.645.5500
Tel: 800.227.8766
Web: www.mksinst.com

万机仪器 (上海) 有限公司  MKS Instruments, United Kindom    MKS Instruments Deutschland GmbH  日本エム・ケー・エス株式会社 MKS Instruments  MKS France  MKS Instruments, Korea  MKS Denmark ApS  español MKS Logo Shop     |      privacy policy     |      legal     |      site map

mksinst.mobi - MKS Instruments for smart phones.